MAUDE Adverse Event Report: CORDIS CASHEL SABER 8MM2CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL SABER 8MM2CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48008002S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with this lot presented no issues during the manufacturing process that can be related to the reported complaint.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the balloon of 8 mm 2 cm 90 saber percutaneous transluminal angioplasty (pta) was used but it ruptured within its nominal pressure.There was no reported patient injury.The device will not be returned for evaluation.
 
Additional Manufacturer Narrative
Complaint conclusion: the balloon of an 8mm x 2cm 90cm saber percutaneous transluminal angioplasty (pta) balloon catheter was used but it ruptured within its nominal pressure.There was no reported patient injury.Multiple attempts were made to obtain more information without success.The product was not returned for analysis.A product history record (phr) review of lot 17718394 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/ below rbp¿ could not be confirmed as the product was not returned for analysis.The exact cause could not be conclusively determined.Vessel characteristics or procedural factors may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium.If resistance is felt upon removal, then the balloon, guidewire and the sheath should be removed together as a unit, particularly if balloon rupture or leakage is known or suspected.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.Forceful handling can result in catheter separation and the subsequent need to use a snare or other medical interventional techniques to retrieve the pieces.Always verify integrity of the catheter after removal.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 8MM2CM 90
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI 
MDR Report Key9466728
Report Number9616099-2019-03403
Device Sequence Number16680911
Product Code LIT
UDI-Device Identifier20705032068885
UDI-Public20705032068885
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model Number48008002S
Device Catalogue Number48008002S
Device Lot Number17718394
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/14/2019
Supplement Date Received by Manufacturer12/16/2019
Initial Report FDA Received Date12/13/2019
Supplement Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-