• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY HERCULES 3 STAGE BALLOON ESOPHAGEAL; KNQ, DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY HERCULES 3 STAGE BALLOON ESOPHAGEAL; KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Model Number G48732
Medical Device Problem Codes Burst Container or Vessel (1074); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.A functional test was performed on the returned device; a cook dilation syringe was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was attempted to be inflated with water.The balloon would not inflate and water was observed leaking from a rupture on the distal side of the balloon.A visual examination of the catheter did not show any kinks/bends.No portion of the device appeared to be missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.However, the device history record for the lot number said to be involved does contain a nonconformance that could potentially be related to the reported observation.The device goes through different inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual product handling conditions could not be duplicated in the laboratory setting.This limits our ability to conclusively determine a cause.A rupture or tear in the balloon can occur if the balloon material comes into contact with a sharp object.Another contributing factor to a rupture in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage balloons esophageal are subjected to a leak test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an esophageal dilation procedure, the technician used a cook hercules 3 stage balloon esophageal.The balloon would not hold pressure.The technician noted it appeared there was some water leaking out the distal end of the balloon.A new cook hercules balloon was opened and used to complete the procedure successfully.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE BALLOON ESOPHAGEAL
Common Device Name
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key9468314
Report Number1037905-2019-00767
Device Sequence Number14452891
Product Code KNQ
UDI-Device Identifier00827002487321
UDI-Public(01)00827002487321(17)220920(10)W4263858
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2022
Device Model NumberG48732
Device Catalogue NumberHBD-15-16.5-18
Device Lot NumberW4263858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2019
Initial Report FDA Received Date12/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-