| Model Number |
Z9002 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a z9002 intraocular lens (iol) was explanted from the patient's eye.Reason for explant is unknown/not provided.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received through product visual testing revealed the lens haptics were observed distorted related to the handling/preparing the unit for surgery.There is no indication that the lens was explanted from the patient's eye during secondary procedure.Therefore, a correction to the initial reporting of an explant is as follows: (b)(4).Device available for evaluation; returned to manufacturer on: 1/29/2020.Device evaluation: the product was returned to the manufacturing site in its original package.Visual inspection using magnification was performed to the returned sample: loose fibers/particles were observed on lens related the handling of the unit out of a sterile environment.Residues of lubricant material was also observed on lens.Both haptics was observed distorted.No damaged was observed to the lens.The condition in which the sample returned is consistent with a lens that was handled and prepare for surgical process.There is no specific failure against the lens reported.Therefore, it is not known that it will be verified.Based on the analyzed of the returned, there is no indication of product quality deficiency.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no additional investigation requests for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|