ADVANCED STERILIZATION PRODUCTS STERRAD® 100NX STERILIZER; STERRAD EQUIPMENT Back to Search Results
Catalog Number 10104008
Medical Device Problem Code Device Emits Odor (1425)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A field service engineer was dispatched to the customer site.The catalytic converter was loose and was adjusted/tightened to resolve the odor/smells issue.Unit meets specifications and was returned to service.Initial reporter phone #: (b)(6).Asp complaint ref #: (b)(4).
 
Event or Problem Description
A customer reported an event of an odor or smell emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
Asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the odor/smells issue, system risk analysis (sra) and functional analysis.The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the odor/smells issue was reviewed within the past six months and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." no parts were replaced as a result of this issue.The assignable cause of the odor/smells issue is the catalytic converter.The field service engineer adjusted/tightened this part and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® 100NX STERILIZER
Common Device Name
STERRAD EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
MDR Report Key9469553
Report Number2084725-2019-01014
Device Sequence Number3987429
Product Code MLR
UDI-Device Identifier10705037048700
UDI-Public10705037048700
Combination Product (Y/N)N
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number10104008
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/25/2019
Supplement Date Received by Manufacturer01/07/2020
Initial Report FDA Received Date12/13/2019
Supplement Report FDA Received Date01/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1026-1027-2013
Patient Sequence Number1
-
-