| Model Number |
500-56140 |
| Medical Device Problem Code |
Audible Prompt/Feedback Problem (4020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to the manufacturer and investigations on 09 december 2019 showed a short in an msd element that appeared to cause a burn in the coolant shaft.The burn did not completely melt through the lumen and there was no communication between any of the drug lumens and the coolant lumen.The device short was not reported by the user.No definitive root cause could be identified.Review of the device history record confirmed that the product was manufactured per standard processes and met all acceptance criteria.There have been no other reported complaints from msds of the same lot.No patient harm was reported.The customer stated the temperature alarm occurred on the third day of treatment.The ifu states ""cumulative ultrasound therapy time should be limited to 24 hours per device." the ifu warns the user to not advance if resistance is met.Excessive force against resistance may result in damage to the device or vasculature.User error was not confirmed but could not be ruled out as a contributing factor to the msd short.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, ekos distributor reported an event that on the third day of treatment, temperature too high sign was noticed on the control unit (cu).The customer flushed the catheter per hospital's protocol but could not resolve the alarm.The sign was still showing.No other information was provided.There were no patient consequences reported.The device was returned to the manufacturer and investigations on 09 december 2019 showed a short in an microsonic device (msd) element that appeared to cause a burn in the coolant shaft.This was not reported by the customer.
|
| |
|
Search Alerts/Recalls
|