| Catalog Number |
115-469 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side rupture.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified broken shell, missing a piece of shell (0-25%), and crease flat.Weight measurement of the device identified device was underweight.A microscopic analysis was performed which identified broken sharp edge on the radius.A dimensional measurement in the shell was performed which identified the thickness was within specification.Based on the device analysis the final assessment was a sharp broken on edge the radius assessed as unidentified (tear) opening, and missing shell due to inconclusive.
|
| |
|