Model Number FG15150-0615-1S
Medical Device Problem Codes Difficult or Delayed Activation (2577); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 11/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned; the reported event could not be confirmed.The cause of the event could not be conclusively determined from the reported information.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report of a perforation during a procedure.The patient was undergoing embolization treatment of an aneurysm.Vessel tortuosity was described as normal.The devices were prepared as indicated in the ifu.It was reported that during the procedure, the phenom felt very stiff and as a result, the wire was navigated further distally, which caused perforation.The perforation was treated by sacrificing the vessel with coils.Post-procedure, the patient was hemiplegic and could not sit up longer than 15 seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATHERA
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key9470358
Report Number2029214-2019-01230
Device Sequence Number17176148
Product Code KRA
UDI-Device Identifier00814584028981
UDI-Public00814584028981
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K151638
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model NumberFG15150-0615-1S
Device Lot NumberJA19-042
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2019
Initial Report FDA Received Date12/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-