DAVIS & GECK CARIBE LTD CHROMIC GUT; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Model Number CG-811
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the suture thread was broken.Another suture was opened and used to complete the case.There was no patient injury.
 
Event or Problem Description
According to the reporter, during a cesarean section procedure , the suture thread was broken.Another suture was opened and used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: a2, a3, b5, d10, e1(prefix, last name), e3, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of sixteen devices, fifteen of which were sealed representative samples.Four open packages were also received.A visual inspection of the opened product noted five foil pouches and retainers as well as three breathable pouches.One open suture and needle were noted.Microscopic evaluation of the returned suture noted no signs of breakage.A tactile and microscopic inspection of the sealed sutures was performed with no abnormalities.The foils were inspected with light assisted microscopy with no abnormalities.A tensile test was performed on the sealed samples and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMIC GUT
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9472841
Report Number9612501-2019-02442
Device Sequence Number2763916
Product Code GAL
UDI-Device Identifier10884521053762
UDI-Public10884521053762
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K885018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model NumberCG-811
Device Catalogue NumberCG-811
Device Lot NumberD8L1148ZY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2020
Initial Date Received by Manufacturer 12/03/2019
Supplement Date Received by Manufacturer04/03/2020
Initial Report FDA Received Date12/16/2019
Supplement Report FDA Received Date04/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
-
-