| Model Number |
STRIP, RLN TMX 100CTMG/DL |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Most likely underlying root cause : mlc-20: user's test strip had poor storage.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results obtained of 196 mg/dl.The customer¿s expected fasting blood glucose test result range is 90 - 125 mg/dl.Customer stated that he only tests in the morning fasting, stated he has never tested fasting in the evenings.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a back to back blood test was not performed by the customer.The product is not stored according to specification and it is stored in the bathroom.The test strip lot manufacturer¿s expiration date is 02/21/2021.Customer did not have another vial of test strips that had been stored and handled correctly.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 07-may-2020: h6: updated fda's method, result, and conclusion codes.H10: meter was not returned for evaluation.Test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.
|
| |
|
Search Alerts/Recalls
|