• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22351C
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of bladder tumour (tur-bt) procedure, the roller ball at the distal end of the hf resection electrode broke off and fell into the patient¿s bladder.The broken off ball was reportedly retrieved but it is unclear whether fragments of the wire remained inside the patient.The intended procedure was then completed using a similar device and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for investigation but to olympus medical systems corporation (omsc), japan (returned to omsc on 2019-12-12).The investigation confirmed that the roller ball at the distal end of the hf-resection electrode is broken off.There are burn marks both on the roller ball and the resection electrode¿s blue insulation tube.The fork tubes are no longer parallel and the electrode is bent at the proximal end.The type of damage found indicates that excessive force must have been applied to the resection electrode¿s distal end.As reported by the customer, the electrode was alternately used for coagulation and ablation during the procedure, which indicates intensified use of the resection electrode thus, this event/incident was attributed to use error.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf-resection electrode without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE
Common Device Name
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key9473568
Report Number9610773-2019-00179
Device Sequence Number2751697
Product Code FAS
UDI-Device Identifier14042761051719
UDI-Public14042761051719
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/27/2023
Device Model NumberWA22351C
Device Catalogue NumberWA22351C
Device Lot Number1000036287
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/06/2019
Supplement Date Received by Manufacturer12/27/2019
Initial Report FDA Received Date12/16/2019
Supplement Report FDA Received Date01/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-