ARTHREX, INC. UKA TIBIAL BEARING, SIZE 1, 8MM, VIT-E; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number UKA TIBIAL BEARING, SIZE 1, 8MM, VIT-E
Health Effect - Clinical Code Pain (1994)
Date of Event 11/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.It was reported the risk manager will not release the device.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
The patient had original surgery for a uni knee replacement on (b)(6) 2018.She reported pain and instability; xrays showed that the poly had worn down.Arthrex parts ar-521-tba8, ar-501-ufld and ar-501-ttla were removed.A total knee revision was done on (b)(6) 2019 with a competitor's device.Patient female, 60s.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UKA TIBIAL BEARING, SIZE 1, 8MM, VIT-E
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key9473603
Report Number1220246-2019-01493
Device Sequence Number16856954
Product Code HSX
UDI-Device Identifier00888867231283
UDI-Public00888867231283
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K161060
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUKA TIBIAL BEARING, SIZE 1, 8MM, VIT-E
Device Catalogue NumberAR-521-TBA8
Device Lot Number5791622
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2019
Initial Report FDA Received Date12/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-