Patient information not provided.Exact date of event is unknown.Event was reported to occur on approx.(b)(6) 2019.Lot # not provided, therefore expiration date is unknown.Lot # not provided, therefore manufacture date is unknown.This complaint is for a flammability event that occurred in (b)(6) 2019.The flammability event is not due to a product defect or problem with this specific lot.The product is flammable until it is completely dry and vapors have dissipated.A flammability event will occur if the product is exposed to an ignition source before the vapors are dissipated.There is flammability information in cavilon no sting barrier film ifus and a flame symbol on each individual pouch.For this application, cavilon no sting barrier film catalog number 3344ens is provided non-sterile to acelity as a component of their kit.I it is unclear if the cavilon¿ no sting barrier film flammability symbol could be seen in the kit.Cautery was the ignition source for this event.The root cause of the flammability event is cautery ignited cavilon no sting barrier film vapors before the vapors dissipated.End of report.
|