| Model Number |
MS1-5590S |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/05/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation nor were x-rays or ultrasound images provided to confirm the alleged event.No root cause can be confirmed at this time.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed on (b)(6) 2019.As per reporter, the surgeon observed loose screws (unknown product) at the lower end of the construct and decided to revise the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation findings: the rod was received for visual inspection as the device was not returned for a mechanical issue.Visual inspection revealed that the returned rod was partially distracted with score marks due to incremental distraction.There was also debris on the distraction rod, which confirmed the reported failure mode.Review of the device history record revealed no discrepancies when the rod was manufactured.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
|
| |
|
Event or Problem Description
|
|
This report has been updated to include investigation findings.
|
| |
|