MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133FV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133FV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133FV-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/31/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: yellow particles and opening on radius.Leak test and microscopic analysis was performed which identified: striated opening.Based on the device analysis the final assessment is: a striated opening due to surgical damage consist in the use of some surgical tool.Reason for reoperation: deflation.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
 
Event or Problem Description
Healthcare professional reported "the expander replaced had deflated during chemotherapy." unknown side.Device explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133FV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9474977
Report Number9617229-2019-13959
Device Sequence Number19508345
Product Code LCJ
UDI-Device Identifier10888628004269
UDI-Public10888628004269
Combination Product (Y/N)N
Initial Reporter StateND
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/08/2022
Device Catalogue Number133FV-15-T
Device Lot Number3163027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2019
Initial Report FDA Received Date12/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-