| Catalog Number |
133FV-15-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: yellow particles and opening on radius.Leak test and microscopic analysis was performed which identified: striated opening.Based on the device analysis the final assessment is: a striated opening due to surgical damage consist in the use of some surgical tool.Reason for reoperation: deflation.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "the expander replaced had deflated during chemotherapy." unknown side.Device explanted.
|
| |
|