| Model Number |
97725 |
| Medical Device Problem Codes |
Break (1069); Device Damaged Prior to Receipt by User (1570); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 977d260, serial#: (b)(4), product type: screening device.Other relevant device(s) are: product id: 977d260, serial/lot #: (b)(4), ubd: 05-sep-2023, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider (hcp) and manufacturing representative (rep) regarding a patient's implantable neurostimulator (ins).Information was reported that the lead is bad.Contacts 0-4 are bad.There are no known factors that may have led to this.A new wens was opened and that's when the bad lead was determined to be the issue.A new lead was used and the issue was resolved.All contact 0-4 were red.No further complications were reported.No patient symptoms were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the lead found no anomalies.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was reported that the cause of the bad contacts was unknown.The lead was never implanted in the patient.This was an out of box failure.No further information was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|