| Catalog Number |
SMXP135RH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a caucasian (b)(6) patient underwent a primary breast reconstruction with a 500cc mentor artoura breast tissue expander and a mentor cpx4 with suture tabs medium height smooth expander and experienced postoperative bilateral deflation.As a result, the patient's devices were explanted on (b)(6) 2019.This report is for the patient's mentor artoura breast tissue expander.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2020, the product investigation was completed.The device was not returned; however, mentor received a photograph of the complaint device.Upon visual inspection of the picture, it was observed a rupture in the middle suture tab.No other anomalies were observed.The photo does not provide enough evidence to determine root cause.Hands on analysis should provide the evidence necessary to confirm the root cause.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|