(b)(4).A manufacturing record evaluation was performed for the finished device lot mpr845, and no non-conformances were identified.Investigation summary = > it was reported pull off - suture needle.Four unopened samples of product code m8735, lot mpr845 were returned for analysis.The product code is multi-strand and contains four strands per packet.During visual inspection of four unopened samples, no defects were found on the packages.Samples were opened and contains four strands double armed; the swage and attachment area of four samples were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged were found.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.Representative samples returned to support investigation of 2 device issues captured in reports 2210968-2019-90940 and 2210968-2019-90965.Analysis results have been included in follow up reports for each.
|