MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US HP THREADED PINS HEADED 6 PK; KNEE INSTRUMENT : FIXATION PINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US HP THREADED PINS HEADED 6 PK; KNEE INSTRUMENT : FIXATION PINS Back to Search Results
Model Number 9505-02-089
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 11/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The threaded pin got stuck in the cutting block.One they could take out but quite deformed the other one was stuck completely.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: examination of the returned device confirmed the reported event.The noted damage is consistent with device wear out through heavy high cycle usage in the field and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP THREADED PINS HEADED 6 PK
Common Device Name
KNEE INSTRUMENT : FIXATION PINS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9477002
Report Number1818910-2019-123309
Device Sequence Number19323977
Product Code HSB
UDI-Device Identifier10603295227922
UDI-Public10603295227922
Combination Product (Y/N)N
Initial Reporter CountrySW
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9505-02-089
Device Catalogue Number950502089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Initial Date Received by Manufacturer 11/25/2019
Supplement Date Received by Manufacturer12/30/2019
01/30/2020
Initial Report FDA Received Date12/16/2019
Supplement Report FDA Received Date01/07/2020
02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-