• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA AF MRI S VR SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA AF MRI S VR SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVFC3D1
Medical Device Problem Codes Signal Artifact/Noise (1036); Mechanical Problem (1384); Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that approximately two weeks post an implantable cardioverter defibrillator (icd) changeout the right ventricular (rv) lead had triggered a lead integrity alert (lia) with noise that was able to be reproduced with isometrics and by 'tapping the can'.It was noted that prior to the icd changeout there had not been any noise seen on the device or lead and it was assumed that there was a loose set screw.A revision procedure was conducted, and the pocket was opened.The cathode was completely thru the header, and a tug test showed no issue.Cables were connected and the lead was manipulated with noise being present, so the physician decided to implant a new rv lead and df4 connector icd on the off chance that it was a header issue.The device was extracted and replaced and the lead was partially removed, capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA AF MRI S VR SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9477322
Report Number3004209178-2019-23923
Device Sequence Number16869605
Product Code LWS
UDI-Device Identifier00643169837829
UDI-Public00643169837829
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2020
Device Model NumberDVFC3D1
Device Catalogue NumberDVFC3D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2019
Initial Date Received by Manufacturer 11/26/2019
Supplement Date Received by Manufacturer01/03/2020
Initial Report FDA Received Date12/16/2019
Supplement Report FDA Received Date01/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
-
-