• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24672
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon hole occurred.During preparation of a 4 x 150 x 135 mustang balloon catheter, a hole was noted on the balloon.The procedure was completed with a new mustang balloon catheter.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.:the device was returned for analysis.A circumferential tear was identified within the distal markerband section of the balloon.A red blood like substance was observed in the balloon.A visual and microscopic examination of the balloon material identified no issues that could have contributed to the complaint incident.The rated burst pressure as per the print on the hub is 24 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that that balloon hole occurred.During preparation of a 4 x 150 x 135 mustang balloon catheter, a hole was noted on the balloon.The procedure was completed with a new mustang balloon catheter.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9477798
Report Number2134265-2019-15789
Device Sequence Number572284
Product Code LIT
UDI-Device Identifier08714729793694
UDI-Public08714729793694
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/05/2022
Device Model Number24672
Device Catalogue Number24672
Device Lot Number0024388408
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2020
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer02/13/2020
Initial Report FDA Received Date12/17/2019
Supplement Report FDA Received Date03/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-