COVIDIEN NANOCROSS 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
AB14W025210150 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Remove (1528); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician used nanocross pta balloon catheter during procedure to treat a little calcified plaque lesion in the distal posterior tibial artery (pta) with 30% stenosis.The device was prepped per ifu with no issues identified.During removal of balloon post pta, difficulty removing the balloon following balloon inflation occurred and the device fractured.No portion of the device remain in patient.It was reported that the balloon was pulled apart during removal and both pieces remained on the balloon catheter shaft upon retrieval.The balloon and wire were removed together, both pieces intact.Physician had trouble advancing another wire through the existing sheath, the existing sheath was removed and a new sheath was placed to complete the procedure.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the nanocross that tore apart in the sheath was a 2.0 -2.5 taper.It was discovered that the sheath was the issue.The same access point was used to complete the procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a 0.014" non-medtronic guidewire and a non-medtronic 6frx45 sheath was used.During attempted removal of balloon post pta, difficulty was encountered removing the balloon to the aortic bifurcation.The balloon is reported to have become stuck on the inside of the sheath at the bifurcation.It was reported that the physician forcefully pulled the balloon and wire out of the sheath and the balloon had torn into two pieces.A 0.014" non-medtronic wire was advanced to the posterior tibial artery and a new 3x150 nanocross balloon was used for the balloon retrieval.Upon retrieval, the new balloon became stuck inside the sheath also.The sheath, wire, and nanocross balloon were all pulled together as a unit across the aortic bifurcation.The new nanocross balloon remained fully intact.The physician still experienced resistance at the bifurcation with any device being removing over the bifurcation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the nanocross elite catheter was received to medtronic for evaluation.The device was received within its opened labelled shelf carton, and loosely coiled within its opened labelled pouch.No ancillary device nor cine images were provided for review.The nanocross elite was received in two sections.The proximal section including the y-manifold and proximal balloon cone.The distal section including internal guidewire lumen, four radiopaque marker bands, distal balloon bond, and distal tip.An inventory of the components concluded that all components were accounted for.The balloon chamber exhibited a radial tear and bunched up the material as if the material bunched up at the mouth of the support catheter during withdrawal.The inner guidewire lumen exhibited tensile stretching as a great force was applied to withdraw the bunched-up balloon through the support catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|