• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL INSERTER / EXTRACTOR; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL INSERTER / EXTRACTOR; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the femoral impactor came apart during impaction.The plastic head was also cracked.This had no effect on patient but surgeon required another instrument to complete the procedure.There is no additional information at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned product confirmed femoral inserter and extractor was disassembled and fractured stop pin.Also, the impaction or chisel-like mark was noted in the head of the instrument.The impactor pad was fractured.Impactor pad and femoral inserter and extractor was submitted for further analysis.The analysis of the impactor pad identified the fracture was consistent with the device fractures analyzed in zrm_wa_0507_19, which indicates overload failure or low cycle fatigue culminating in the overload failure.The analysis of the femoral inserter and extractor identified the fracture of the stop pin was suspected due to sheared overload.The material analysis found consistent with the print specification.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL INSERTER / EXTRACTOR
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9480321
Report Number0001822565-2019-05298
Device Sequence Number234230
Product Code MBH
UDI-Device Identifier00889024232860
UDI-Public(01)00889024232860
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42509909200
Device Lot Number63871202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/27/2019
Supplement Date Received by Manufacturer05/26/2020
Initial Report FDA Received Date12/17/2019
Supplement Report FDA Received Date05/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-