| Model Number |
ASD27A |
| Medical Device Problem Code |
Expulsion (2933)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # :p050006.
|
| |
|
Event or Problem Description
|
|
It was reported the physician implanted a 27mm gore® cardioform asd occluder to close an atrial septal defect on (b)(6) 2019.The defect was not balloon sized.Transthoracic echocardiography on (b)(6) 2019 showed the device had embolized to the left pulmonary artery.The embolized device was removed by snare in a transcatheter procedure and a 32mm gore® cardioform asd occluder was implanted.The patient is currently stable and doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional case details: static defect measurement was 10mm; the patient had a deficient aortic rim (>5mm); there was a sufficient posterior rim (no measurements).The gore® cardioform asd occluder instructions for use list embolization as potential clinical and device adverse event.
|
| |
|