• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF XLPE 32 20 DEG 54-56 F; PROSTHESISHIPSMICNSTRANDMTALPLYMRCEMNTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF XLPE 32 20 DEG 54-56 F; PROSTHESISHIPSMICNSTRANDMTALPLYMRCEMNTD Back to Search Results
Model Number 71333335
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/21/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, the reflection liner couldn't be implanted into the cup.Finally a r3 liner and cup were used to complete the surgery.No injury or impact to patient reported.Delay of 1 hour reported.
 
Additional Manufacturer Narrative
The associated devices, used in treatment, were returned and evaluated.A visual inspection of the returned devices could not confirm the stated failure mode.The devices have some signs of damage from implantation / insertion / extraction.The devices were manufactured in 2019.A dimensional inspection was attempted on the liner but it was too damaged from the attempted insertion to obtain accurate measurements.A dimensional inspection of the shell found it to be within specification.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes of the reported event could include improper insertion technique or device damaged during insertion.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF XLPE 32 20 DEG 54-56 F
Common Device Name
PROSTHESISHIPSMICNSTRANDMTALPLYMRCEMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9480950
Report Number1020279-2019-04426
Device Sequence Number209668
Product Code JDI
UDI-Device Identifier03596010455680
UDI-Public03596010455680
Combination Product (Y/N)N
PMA/510(K) Number
K002747
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71333335
Device Catalogue Number71333335
Device Lot Number19HM10458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2020
Initial Date Received by Manufacturer 11/22/2019
Supplement Date Received by Manufacturer11/22/2019
Initial Report FDA Received Date12/17/2019
Supplement Report FDA Received Date04/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-