• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM DRIVE IMPLANT 5.0X13 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM DRIVE IMPLANT 5.0X13 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.998
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Pain (1994); Tissue Damage (2104)
Date of Event 11/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The dentist has exceeded the recommended installation torque (60 ncm).Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system and is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a new follow-up report will be send.
 
Event or Problem Description
(b)(4)¿ the dentist reported that 1 year and 1 month after the dental implant was installed in intraoral region 23#, its non- osseointegration was observed.Moreover, 80n.Cm of primary stability was achieved, the patient presented insufficient bone quality/quantity (bone type ii), immediate implant was performed and immediate load procedure was done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE IMPLANT 5.0X13 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key9481309
Report Number3008261720-2019-06694
Device Sequence Number3541836
Product Code DZE
UDI-Device Identifier07899878023371
UDI-Public07899878023371
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 12/17/2019
Report Date (Section F) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2022
Device Catalogue Number109.998
Device Lot Number800325536I
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/17/2019
Device Age25 MO
Event Location Other
Date Report to Manufacturer11/21/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2019
Initial Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-