• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE
Device Problem Imprecision (1307)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 11/19/2017
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose.Once the evaluation is performed, a follow-up medwatch report will be submitted.(b)(4).
 
Event Description
On (b)(6) 2019 a conference call was held with the surgeon, about the ongoing pmcf for the rosa one device.The surgeon transmitted information about adverse events that he recorded.An incorrect k-wire positioning (too medial) was noted during an intra-operative ct-exam.A per-operative re-positioning was performed.The surgeon detailed that the patient had an instable anatomy related to trauma and patient reference on iliac crest.Spine movements were not transmitted to the iliac crest.Level of spine treated l1.Also, due to dry cough during the first acquisition the registration has been performed two times.According to the gertzbein and robbins classification system, the guidewire accuracy was grade a (0mm) for one trajectory and was grade b (<2mm) for one other trajectory.According to the surgeon the severity is mild.Relation to device: probably // relation to instrument: not related // relation to procedure: definitely.
 
Event Description
On december 02nd, 2019 a conference call was held with the surgeon, about the ongoing pmcf for the rosa one device.The surgeon transmitted information about adverse events that he recorded.An incorrect k-wire positioning (too medial) was noted during an intra-operative ct-exam.A per-operative re-positioning was performed.The surgeon detailed that the patient had an instable anatomy related to trauma and patient reference on iliac crest.Spine movements were not transmitted to the iliac crest.Level of spine treated l1.Also, due to dry cough during the first acquisition the registration has been performed two times.According to the gertzbein and robbins classification system, the guidewire accuracy was grade a (0mm) for one trajectory and was grade b (<2mm) for one other trajectory.According to the surgeon the severity is mild.Relation to device: probably // relation to instrument: not related // relation to procedure: definitely.
 
Manufacturer Narrative
A full analysis of the data logs could not be performed due to that data were never provided despite multiple attempts to retrieve them.The event is non-verifiable and the technical root cause cannot be determined.
 
Event Description
On (b)(6)2019 a conference call was held with the surgeon, about the ongoing pmcf for the rosa one device.The surgeon transmitted information about adverse events that he recorded.An incorrect k-wire positioning (too medial) was noted during an intra-operative ct-exam.A per-operative re-positioning was performed.The surgeon detailed that the patient had an instable anatomy related to trauma and patient reference on iliac crest.Spine movements were not transmitted to the iliac crest.Level of spine treated l1.Also, due to dry cough during the first acquisition the registration has been performed two times.According to the gertzbein and robbins classification system, the guidewire accuracy was grade a (0mm) for one trajectory and was grade b (<2mm) for one other trajectory.According to the surgeon the severity is mild.Relation to device: probably // relation to instrument: not related // relation to procedure: definitely.
 
Manufacturer Narrative
Corrected data: - b1 report type - b4 date of this report - g4 date received by manufacturer - h1 type of reportable event - h2 if follow-up, what type - h6 event problem and evaluation codes - h10 additional narratives/data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
MDR Report Key9482083
MDR Text Key176453083
Report Number3009185973-2019-00455
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K182848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA ONE
Device Catalogue NumberROSAS00201
Device Lot Number3.1.0.1266
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received02/14/2020
07/30/2020
Supplement Dates FDA Received03/12/2020
08/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-