| Model Number |
FL23SE |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was evaluated by service technician of distributor and picture was sent to the manufacturer for analysis.Picture analysis demonstrated that the customer had modified the cord with a pvc pipe that did not allow the cord to flex freely when bed is lowered.
|
| |
|
Event or Problem Description
|
|
It was reported that the bed controls of a fl23se hospital bed were not working.It was further observed that the power cord was torn.There was no patient involvement, no impact or adverse event.
|
| |
|