MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000E
Device Problems Defective Component (2292); Output below Specifications (3004)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the arcticsun device was getting 113 alert. Per troubleshooting, manual control was enabled and the device was set to 4c and then 38c. The device heated and cooled as expected. With only the fluid delivery line attached, the flow rate was 1. 6l/min, the inlet pressure was -7psi, and the circulation pump was 100%. Four out the six valve sets were out of specification. The other two were -6. 9psi. The flow rate fluctuated between 3. 4 and 4. 0l/min. It was later reported via (b)(4), biomed (b)(6) stated that he changed out the valve sets on the unit that were out of specification. They all read within spec after replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9484057
MDR Text Key180339787
Report Number2020394-2019-05659
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-