| Model Number |
980X3SKNISC |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the service engineer (se) evaluated the device and identified relevant error messages in the ventilator¿s memory logs but was unable to duplicate the reported issue.The ventilator passed all testing per manufacturing specification and was placed back into clinical use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, while in use on a patient, a 980 ventilator stopped working.The patient was removed from the ventilator and placed on an alternate ventilator with no harm or injury.
|
| |
|