Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Wheal(s) (2241)
|
Event Type
Injury
|
Manufacturer Narrative
|
The device has not yet been returned.An investigation will be conducted once the sample has been returned and a supplemental report will be submitted.Age: asked, but unknown.Date of birth: asked, but unknown.Sex: asked, but unknown.Weight: asked, but unknown.Ethnicity: asked, but unknown.Race: asked, but unknown.Date of event: asked, but unknown.Model number: not applicable.Catalog number: not applicable.Lot number: not applicable.Expiration date: not applicable.Udi: not available.
|
|
Event Description
|
It was reported that a patient experienced an allergic reaction after using an xtra silent nite slide link appliance.According to the information provided by the doctor, the patient experienced hives on unknown date.Upon experiencing the reaction, the patient removed the device and the symptoms went away.The patient tried to use the device once more, but symptoms recurred.The patient does not have any known allergies.
|
|
Manufacturer Narrative
|
The device was not available for evaluation.A batch/lot review for the associated material showed no manufacturing deviations or abnormalities.Glidewell research team and namsa conducted a series of testing on erkodent material (erkoloc-pro and erkodur) following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The test results were listed and summarized in rpt 9733 rev 1.0.For cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.For skin irritation, there was no erythema and no edema observed on the skin test.For sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.The test article showed nonirritant to the oral mucosa as compared to the control article.This complaint will be kept on record for track and trending purposes.
|
|
Search Alerts/Recalls
|
|