| Model Number |
MICL 12.6 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Not Applicable (3189); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight:unk.Ethnicity:unk.Race:unk.Date of event:unk.Device evaluation : lens was returned dry, in lens case.Visual inspection found a haptic torn and bent.Residue on the lens was also observed.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm, micl12.6, - 16.00 diopter implantable collamer lens into the patients right eye (od) on (b)(6) 2019.On (b)(6) 2019 the lens was exchanged for a shorter length lens due to excessive vault.This exchange resolved the problem.Enlargement of pi (yag) was also reported as a secondary surgical intervention.In the reporters opinion the cause of this event was patient related factor.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h6 - patient code: 3189 should be corrected to 3191.Claim#: (b)(4).
|
| |
|