• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL 12.6
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight:unk.Ethnicity:unk.Race:unk.Date of event:unk.Device evaluation : lens was returned dry, in lens case.Visual inspection found a haptic torn and bent.Residue on the lens was also observed.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 12.6mm, micl12.6, - 16.00 diopter implantable collamer lens into the patients right eye (od) on (b)(6) 2019.On (b)(6) 2019 the lens was exchanged for a shorter length lens due to excessive vault.This exchange resolved the problem.Enlargement of pi (yag) was also reported as a secondary surgical intervention.In the reporters opinion the cause of this event was patient related factor.
 
Additional Manufacturer Narrative
Corrected data: h6 - patient code: 3189 should be corrected to 3191.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key9484640
Report Number2023826-2019-02417
Device Sequence Number4202321
Product Code MTA
UDI-Device Identifier00841542103145
UDI-Public00841542103145
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Model NumberMICL 12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2019
Initial Date Received by Manufacturer 11/19/2019
Supplement Date Received by Manufacturer02/03/2020
Initial Report FDA Received Date12/17/2019
Supplement Report FDA Received Date02/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
-
-