Related manufacturer reference number: 2938836-2019-17386; 2938836-2019-17388.It was reported that following the patient¿s death, the device was interrogated and revealed episodes of inappropriate auto-mode switch (ams) due to atrial lead noise.An electrogram (egm) revealed sensing of ventricular tachycardia (vt) and recorded by the device as high ventricular rate (hvr).Some undersensing was observed on the pacemaker.Loss of capture was observed on the atrial and ventricular leads.The cause of the patient¿s death was vt and there was not fault of the leads nor pacemaker.
|