| Model Number |
MMT-723NAL |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.The customer¿s blood glucose was unknown at the time of incident.The customer also stated that the insulin pump had crack on the screen and also had no delivery alarm.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test.No excessive no delivery or occlusion alarm noted.No motor error alarm noted.Motor passed motor test.Device received with cracked case from display window corner and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|