| Model Number |
EEAXL2835 |
| Medical Device Problem Codes |
Device Slipped (1584); Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic sigmoidectomy, while being applied on the sigmoid colon, the space between the cartridge and anvil did not close and was the green bar was not visible in the indicator window.The device did not fire.Another device was used to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the staple guide noted the presence of a full complement of staples.The anvil of the instrument was observed to be not tilted and separated from the instrument.Functionally, the device was applied over the appropriate test media producing acceptable results.The knife cut the test media cleanly and completely and the staple line was properly formed.There was no difficulty in closing the anvil and cartridge.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported conditions.If information is provided in the future, a supplemental report will be issued.
|
| |
|