MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD UD 27IN 4-0 S/A J-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD UD 27IN 4-0 S/A J-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J455H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent an unknown surgery on (b)(6) 2019 and suture was used.During the procedure, the needle was broken at the swage part during dermostitch.It was not a control release needle.The broken piece was retrieved, and no pieces remained in the patient.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 1/17/2020.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 2/19/2020.Additional information: d10, h4, h6.A manufacturing record evaluation was performed for the finished device pck183 batch number, and no non-conformances were identified.Additional h3 investigation summary: a suture needle was returned for evaluation.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A scanning electron microscope (sem) was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations in order to determine the fracture mode.The evaluation of (b)(4) revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile overload failure.Mechanical damage, a cup-and-cone shaped fracture and macroscopic plastic deformation observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD UD 27IN 4-0 S/A J-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9488883
Report Number2210968-2019-91060
Device Sequence Number3541721
Product Code GAM
UDI-Device Identifier10705031040038
UDI-Public10705031040038
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberJ455H
Device Catalogue NumberJ455H
Device Lot NumberPCK183
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Initial Date Received by Manufacturer 11/28/2019
Supplement Date Received by Manufacturer12/22/2019
01/27/2020
Initial Report FDA Received Date12/18/2019
Supplement Report FDA Received Date01/17/2020
02/19/2020
Patient Sequence Number1
-
-