| Model Number |
N/A |
| Medical Device Problem Codes |
Mechanical Problem (1384); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No information was provided.The customer reported the following lot #'s may be have been associated with this incident; 2021-09 eu, 2021-09 et, 202105 fd, 202110 fg & 202109 ey.Reported lot #s expiration date information requested.Waiting on this information.Reported lot #s manufacture date information requested.Waiting on this information.The sample has been received for evaluation.Analysis is pending.A device history review (dhr) was requested for the reported lot #s.A supplemental report will be sent upon sample evaluation completion.
|
| |
|
Event or Problem Description
|
|
A hospital employee reported that 3m¿ red dot¿ monitoring electrodes with foam tape and sticky gel, model 2570-5, are frequently getting stuck in the lead wires after they are used.Phillips¿ monitors and lead wires are used by the facility.No patient injury was alleged.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned complaint samples and plant retains were tested and exceeded the minimum requirements for stud pull testing.No adverse trend was observed for this issue.3m will continue to monitor.
|
| |
|