| Model Number |
TA4535L |
| Medical Device Problem Codes |
Failure to Fire (2610); Expulsion (2933)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, while the nurse was preparing the device during a procedure, the device did not fire because it had prefired.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the single use loading unit (sulu) noted that all of the staples were partially advanced in the cartridge.The staples were in usable unformed condition.Functionally, since the clinical instrument was not received a pmv representative instrument was used for testing.The advanced staples in the sulu were reused and reset inside the cartridge.The representative instrument and sulu were applied to test media with proper staple formation.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the advanced staple condition may occur when the firing stroke is not completed during application.The instructions for use of this product state: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the instrument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The current packaging design for the product does not allow room for advanced staples, thus the condition could not have occurred during assembly or shipment.The root cause of the observed damage was due to the product not being used as indicated which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|