Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Chemical Exposure (2570)
Date of Event 11/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The facility reported a leak of rapicide pa high level disinfectant from their dsd edge automated endoscope reprocessor causing two employees to be sent to the emergency room due to exposure symptoms.Medivators field service engineer (fse) was dispatched to the facility and inspected the aer.The fse found that a manifold connector was not working properly and replaced the part.Test cycles were performed, and it was confirmed the unit was running according to specification.Medivators regulatory followed up with the facility and they confirmed one of the employees was wearing a mask at the time of the incident.One of the employees has asthma and was treated for symptoms in the er.They were both provided a doctor's note to be excused from work for three days.There have been no reports of further incidents.The two employees current condition are reported to be fine.This will continue being monitored in the medivators complaint handling system.
 
Event or Problem Description
The facility reported a leak of rapicide pa high level disinfectant from their dsd edge automated endoscope reprocessor causing two employees to be sent to the emergency room due to exposure symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DSD EDGE
Common Device Name
HIGH LEVEL DISINFECTANT
Manufacturer (Section D)
MEDIVATORS INC.
14605 287th ave n
plymouth MN 55447
MDR Report Key9490209
Report Number2150060-2019-00073
Device Sequence Number1215274
Product Code FEB
UDI-Device Identifier00677964046674
UDI-Public00677964046674
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K092387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Remedial Action Patient Monitoring
Type of Report Initial
Report Date (Section B) 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2019
Initial Report FDA Received Date12/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-