MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that error icon displayed occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that error icon displayed occurred.The product was evaluated.An exterior physical visual inspection was performed and passed.Receiver charged and boot test was performed and passed.Functional test was performed and passed.Open receiver case for internal visual inspection and passed.The log was downloaded and reviewed finding error related to the complaint.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9490442
MDR Text Key175213302
Report Number3004753838-2019-109743
Device Sequence Number1
Product Code MDS
UDI-Device Identifier30386270000157
UDI-Public30386270000157
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTR-DR-001
Device Lot Number5261391
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received12/18/2019
Supplement Dates Manufacturer Received03/22/2020
Supplement Dates FDA Received03/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient Weight79
-
-