MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-040
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Adhesion(s) (1695); Cyst(s) (1800); Edema (1820); Fatigue (1849); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Tissue Damage (2104); Distress (2329); Discomfort (2330); Impaired Healing (2378); Ambulation Difficulties (2544); No Code Available (3191); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Event Date 11/01/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4). Initial reporter occupation: lawyer. (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Medical records received. After review of medical records, the patient was revised to address aseptic loosening, acetabular protusio, pseudotumor secondary to metal on metal reaction, rupture of adductor tendon mechanism and fascial dehiscence. Prior to being revised, patient reports pain. Workup elsewhere reveals a loose cup with acetabular protusio and rotation of the cup into a passive vertical position. Intraoperatively, a cystic mass was opened and there was milky, non-purulent fluid consistent with metal on metal disease. The abductors were repaired. It was confirmed that the acetabular component was loose and had rotated 90 degrees into the base of the acetabulum. Metal on metal metallic debris was found in the posterior acetabular wall as well as on the trunnion. Doi: (b)(6) 2009; dor: (b)(6) 2019; (right hip).
 
Manufacturer Narrative
Product complaint # (b)(4). Investigation summary
==
> no device associated with this report was received for examination. The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system. Depuy considers the investigation closed. Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary. If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTRI-LOCK BPS SZ 4 STD OFFSET
Type of DeviceTRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopedics drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9490548
MDR Text Key179264976
Report Number1818910-2019-123845
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number1012-04-040
Device Catalogue Number101204040
Device Lot NumberD4GFA1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/18/2019 Patient Sequence Number: 1
-
-