• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 86248
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported that failure to halt ablation occurred.A maestro 4000 controller was selected for a ablation procedure.During the ablation it was noticed that the radiofrequency energy (rf) continued after the foot-pedal was released.The issue was corrected by pressing the footswitch again.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO 4000 CONTROLLER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9490585
Report Number2134265-2019-15943
Device Sequence Number2981633
Product Code LPB
UDI-Device Identifier08714729861911
UDI-Public08714729861911
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86248
Device Catalogue Number86248
Device Lot Number0010515004
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/26/2019
Initial Report FDA Received Date12/18/2019
Patient Sequence Number1
-
-