| Model Number |
86248 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that failure to halt ablation occurred.A maestro 4000 controller was selected for a ablation procedure.During the ablation it was noticed that the radiofrequency energy (rf) continued after the foot-pedal was released.The issue was corrected by pressing the footswitch again.No patient consequences were reported.
|
| |
|