| Model Number |
H700150 |
| Medical Device Problem Code |
Failure to Analyze Signal (1539)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation ablation procedure, a cancellation occurred due to intracardiac signal loss.Troubleshooting consisted of bypassing cardo, power cycling, swapping the ecg cables, checking the connection between the stimulator and amplifier, repositioning the right lead, exchanging the right leg lead patch and resetting the notch filter but the issue remained.The procedure was cancelled and there were no adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information received, the cause of the reported procedure cancellation could not be conclusively determined.
|
| |
|