• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700150
Medical Device Problem Code Failure to Analyze Signal (1539)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
During an atrial fibrillation ablation procedure, a cancellation occurred due to intracardiac signal loss.Troubleshooting consisted of bypassing cardo, power cycling, swapping the ecg cables, checking the connection between the stimulator and amplifier, repositioning the right lead, exchanging the right leg lead patch and resetting the notch filter but the issue remained.The procedure was cancelled and there were no adverse patient consequences.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information received, the cause of the reported procedure cancellation could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key9491397
Report Number2184149-2019-00251
Device Sequence Number11648271
Product Code DQK
UDI-Device Identifier05415067001245
UDI-Public05415067001245
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberH700150
Device Catalogue NumberH700124
Device Lot Number6018016
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/02/2019
Supplement Date Received by Manufacturer01/22/2020
Initial Report FDA Received Date12/18/2019
Supplement Report FDA Received Date01/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-