| Model Number |
0134L18 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter balloon burst and fell out of the patient some time after placement.Per additional information from the ibc via email 02dec2019, there were no missing pieces to the burst balloon.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter balloon burst and fell out of the patient some time after placement.Per additional information from the ibc via email 02dec2019, there were no missing pieces to the burst balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed, however, the cause is unknown.Evaluation found an irregular burst and it was noted that it was burst along the inflation eye with no missing pieces.The sample was evaluated under a microscope and no conditions were found that could be associated with the reported event.A potential root cause for this failure mode could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ operator error.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "1.Method for use: (1) do not inflate the balloon in the urethra.(the urethra may be injured) (2) do not pull the catheter hard.(the bladder/urethra maybe injured) 2.Applicable patients (1) patients with delirium who might pull out catheter (when patient tugs at catheter unconsciously, the bladder and urethra may be damaged).Contraindications 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (4) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1) do not use on patients who are or have been allergic to natural rubber latex." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|