The reported event was confirmed.An orange lubricath coude catheter was returned without its original packaging.There were white markings on the catheter balloon and tip area which appear to be indicative of a degradation.A 10cc leur lock syringe was used to inflate the catheter with 10 ccs of air.The balloon appeared to be asymmetric.The catheter was then passively deflated with no issues.Then the catheter was inflated again with 10 ccs of water.The catheter's balloon was out specification as the asymmetry of the balloon when inflated with water and when inflated with air was greater than 60/40.After a few minutes, water began to slowly leak out of the inflation valve.The remaining water was able to deflate with no issues.A potential root cause for this failure could be "valve does not close fully".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿peel to open c.R.Bard, inc.Covington, ga 30014 www.Bardmedical.Com 1-800-526-4455 bardex® foley catheter caution: this product contains natural rubber latex which may cause allergic reactions.Pd-50201 10/00 sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Bard and bardex are registered trademarks of c.R.Bard, inc.Or an affiliate.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|