| Model Number |
CBL-2000 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing date 5/2017.A review of records related to the device including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
| |
|
Event or Problem Description
|
|
Account reported that cable had burn marks on it.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: 1 sealed cable was received within original packaging - reported lot number confirmed.The products were returned within their original box that was opened.A visual inspection was performed on the returned product.The visual inspection of the product reveals brown/dark grey discoloration along the length of the cable.How and when these discolorations were introduced cannot be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|