| Catalog Number |
K6412 SMN 10445104 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mdrs 2517506-2019-00477, 2517506-2019-00479, 2517506-2019-00480, 2517506-2019-00481 were filed for the same event.The customer contacted the siemens customer care center (ccc) and reported an elevated thyroid stimulating hormone (tsh) patient result was obtained on a dimension vista 1500 instrument.Siemens is investigating the event.
|
| |
|
Event or Problem Description
|
|
A discordant, falsely elevated thyroid stimulating hormone (tsh) result was obtained on a patient sample on a dimension vista 1500 instrument.The discordant result was reported to the physician(s) and questioned.The same sample was reprocessed at an alternate facility on a non-siemens alternate methodology and on an alternate siemens methodology and a lower correct result was obtained.There are no known reports of adverse health consequences due to the discordant elevated tsh result.
|
| |
|
Additional Manufacturer Narrative
|
|
Mdr 2517506-2019-00478 was filed on 18-dec-2019.Mdrs 2517506-2019-00477, 2517506-2019-00479, 2517506-2019-00480 and 2517506-2019-00481 were also filed for the same event on 18-dec-2019.Additional information (19-dec-2019): siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant falsely elevated thyroid stimulating hormone (tsh) result.Hsc evaluated the information provided and the instrument data files.There was no indication of instrument issues when the patient's samples were processed.The results of the heterophilic antibody blocking (hab) treatment on the patient sample indicated a difference in recovery indicative of a sample specific interferent.As stated in the instructions for use (ifu) for loci thyroid stimulating hormone (tsh) on the dimension vista; "patient samples may contain heterophilic antibodies that could react in immunoassays to give a falsely elevated or depressed result.This assay has been designed to minimize interference from heterophilic antibodies.Nevertheless, complete elimination of this interference from all patient specimens cannot be guaranteed." the cause of the discordant tsh result is due to a sample specific interferent.The patient sample was not requested to be returned for evaluation because the customer was able to perform adequate testing to determine the cause.This event is singular in nature and is limited to a single patient sample.The instrument is operational.No product non-conformance identified.The device is performing within specifications.No further evaluation is required.Section h6 has been updated to reflect the hsc investigation.
|
| |
|
Search Alerts/Recalls
|