Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.Device evaluation: analysis of the returned device identified: weight to the spec, crease fold, deformation.Cloudy in the gel observed after autoclave cycle.Base on the device analysis the final assessment is:no issues found related with the manufacturing process.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The reason for reoperation is rupture.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|