ENCORE MEDICAL L.P. RSP; RSP MONOBLOCK HEMI ADAPTER W/RETAINING SCREW Back to Search Results
Model Number 510-99-000
Health Effect - Clinical Codes Failure of Implant (1924); Loss of Range of Motion (2032); Discomfort (2330)
Date of Event 11/22/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as patient's discomfort and loss of motion.The previous surgery and the surgery detailed in this event occurred 1 year and 1 month apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to patient's discomfort and loss of motion.There were no findings during this evaluation that indicate the reported devices were defective.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are many factors that may contribute to the event that are outside the control of djo surgical are inadequate soft tissue support, patient age, patient bone deterioration, degenerative bone disease, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient complained of discomfort and loss of range of motion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Common Device Name
RSP MONOBLOCK HEMI ADAPTER W/RETAINING SCREW
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key9495227
Report Number1644408-2019-01222
Device Sequence Number1541825
Product Code PHX
UDI-Device Identifier00888912144773
UDI-Public(01)00888912144773
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111735
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/17/2021
Device Model Number510-99-000
Device Catalogue Number510-99-000
Device Lot Number639G1077
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2019
Initial Report FDA Received Date12/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-