The investigation determined that non-reproducible vitros phyt results were obtained from a single patient sample tested on a vitros 5600 integrated system.The investigation was unable to determine a definitive assignable cause for this event.A sample matrix issue related to the affected patient cannot be ruled out.The issue is isolated only to the single affected patient sample, as other patient samples run over this time frame did not demonstrate a similar issue.Quality control and precision test results from the time frame of the event demonstrated acceptable within-laboratory and within-run performance of the vitros phyt slide lot 2619-0171-2451.With the exception of the non-reproducible results obtained from the affected patient sample, vitros phyt lot 2619-0171-2451 otherwise appeared to be performing within expectations.Additionally, ongoing tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros phyt lot 2619-0171-2451.In addition, results of precision tests were within acceptable guidelines.The vitros 5600 integrated system was operating as expected at the time of the event.Finally, the affected sample was not centrifuged according to the collection device manufacturer¿s recommendations.Therefore, improper pre-analytical sample handling cannot be ruled out as a contributing factor.It is likely that cellular debris, due to poor sample preparation, was present in the affected samples, although this could not be confirmed.
|