| Model Number |
305604 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.The manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that sharps coll slide close chemo 9gal was missing the lid.This was discovered before use.The following information was provided by the initial reporter: material no.: 305604 batch no.: unknown.It was reported a case was received without lids.Per email: we have received a quality complaint on product sold to our customer.Please, respond accordingly with the rma number, if necessary, and the disposition of the customer's concern.Please contact the customer and cc me on any future communication.Injuries or adverse event: no are any samples available for return? yes reported issue: customer receiving 1 cs without lids.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary the actual product nor photo representation was provided.A review of the device history record (dhr) was not possible since a lot number could not be provided.Also, a review of non-conforming material report (ncmr) for the reported container was performed for the past 12 months and no manufacturing or material defects were identified that could have caused or contributed to the reported incident, ¿missing lids¿.As corrective action a weighing system has already been implemented for this container manufacturing process to ensure the correct quantity of pieces per box before shipping.Unfortunately, a weighted label placed on the box by the weighing system is required to identify or confirm this issue.The root cause is unknown.The issue could have occurred during distribution if repackaged by a distributor.Based on these findings, our investigation is inconclusive.Evidence of original packaging is required.Bd will continue to monitor for any trends.H3 other text : see section h.10.
|
| |
|
Event or Problem Description
|
|
It was reported that sharps coll slide close chemo 9gal was missing the lid.This was discovered before use.The following information was provided by the initial reporter: material no.: 305604 batch no.: unknown it was reported a case was received without lids.Per email: we have received a quality complaint on product sold to our customer.Please, respond accordingly with the rma number, if necessary, and the disposition of the customer's concern.Please contact the customer and cc me on any future communication.Injuries or adverse event: no are any samples available for return? yes reported issue: customer receiving 1 cs without lids.
|
| |
|
Search Alerts/Recalls
|